Storing of Extracts in Polypropylene Microcentrifuge Tubes Yields Contaminant Peak During Ultra-flow Liquid Chromatographic Analysis
Parthraj R. Kshirsagar2, Harsha Hegde2,3, Sandeep R. Pai2★★ Corresponding author
- 1Plant Biotechnology and Tissue Culture Division, Regional Medical Research Centre, Indian Council of Medical Research, India.
- 2Traditional and Herbal Medicine Division, Regional Medical Research Centre, Indian Council of Medical Research, Belagavi, Karnataka, India.
CORRESPONDENCE
Sandeep R. Pai
Traditional and Herbal Medicine Division, Regional Medical Research Centre, Indian Council of Medical Research, Belagavi, Karnataka, India.
Received: 12-01-2016; Revised: 10-02-2016.
Volume 12, Issue 46s · pp. S303–S306 · PUBLISHED 7 July 2016 · DOI: 10.4103/0973-1296.185719
View on Pharmacogn. Mag. original site ↗
ABSTRACT
Aim: This study was designed to understand the effect of storage in polypropylene microcentrifuge tubes and glass vials during ultra- flow liquid chromatographic (UFLC) analysis. Materials and Methods: One ml of methanol was placed in polypropylene microcentrifuge tubes (PP material, Autoclavable) and glass vials (Borosilicate) separately for 1, 2, 4, 8, 10, 20, 40, and 80 days intervals stored at −4°C. Results: Contaminant peak was detected in methanol stored in polypropylene microcentrifuge tubes using UFLC analysis. The contaminant peak detected was prominent, sharp detectable at 9.176 ± 0.138 min on a Waters 250–4.6 mm, 4 μ, Nova-Pak C18 column with mobile phase consisting of methanol:water (70:30). Conclusion: It was evident from the study that long-term storage of biological samples prepared using methanol in polypropylene microcentrifuge tubes produce contaminant peak. Further, this may mislead in future reporting an unnatural compound by researchers.
KEYWORDS
REFERENCES
As publishedShowing references and in-text citations exactly as published.
Cite this article
SELECT FORMAT
Kshirsagar, P. R., Hegde, H., & Pai, S. R. (2016). Storing of Extracts in Polypropylene Microcentrifuge Tubes Yields Contaminant Peak During Ultra-flow Liquid Chromatographic Analysis. Pharmacognosy Magazine, 12(46s), S303–S306. https://doi.org/10.4103/0973-1296.185719