Phcog.Net logo

BROWSE ALL JOURNALS

    SEE ALL 6 JOURNALS
    Article

    Storing of Extracts in Polypropylene Microcentrifuge Tubes Yields Contaminant Peak During Ultra-flow Liquid Chromatographic Analysis

    Parthraj R. Kshirsagar2, Harsha Hegde2,3, Sandeep R. Pai2 Corresponding author

    1. 1Plant Biotechnology and Tissue Culture Division, Regional Medical Research Centre, Indian Council of Medical Research, India.
    2. 2Traditional and Herbal Medicine Division, Regional Medical Research Centre, Indian Council of Medical Research, Belagavi, Karnataka, India.

    CORRESPONDENCE

    Sandeep R. Pai

    Traditional and Herbal Medicine Division, Regional Medical Research Centre, Indian Council of Medical Research, Belagavi, Karnataka, India.

    drpaisr@gmail.com

    Received: 12-01-2016; Revised: 10-02-2016.

    Volume 12, Issue 46s · pp. S303–S306 · PUBLISHED 7 July 2016 · DOI: 10.4103/0973-1296.185719

    View on Pharmacogn. Mag. original site ↗

    ABSTRACT

    Aim: This study was designed to understand the effect of storage in polypropylene microcentrifuge tubes and glass vials during ultra- flow liquid chromatographic (UFLC) analysis. Materials and Methods: One ml of methanol was placed in polypropylene microcentrifuge tubes (PP material, Autoclavable) and glass vials (Borosilicate) separately for 1, 2, 4, 8, 10, 20, 40, and 80 days intervals stored at −4°C. Results: Contaminant peak was detected in methanol stored in polypropylene microcentrifuge tubes using UFLC analysis. The contaminant peak detected was prominent, sharp detectable at 9.176 ± 0.138 min on a Waters 250–4.6 mm, 4 μ, Nova-Pak C18 column with mobile phase consisting of methanol:water (70:30). Conclusion: It was evident from the study that long-term storage of biological samples prepared using methanol in polypropylene microcentrifuge tubes produce contaminant peak. Further, this may mislead in future reporting an unnatural compound by researchers.

    KEYWORDS

    Open in new tab

    REFERENCES

    As published

    Showing references and in-text citations exactly as published.

      Cite this article

      SELECT FORMAT

      Kshirsagar, P. R., Hegde, H., & Pai, S. R. (2016). Storing of Extracts in Polypropylene Microcentrifuge Tubes Yields Contaminant Peak During Ultra-flow Liquid Chromatographic Analysis. Pharmacognosy Magazine, 12(46s), S303–S306. https://doi.org/10.4103/0973-1296.185719